    Use of haemocytometer to quantify concentration of cells' suspensions

    abotero

    Published: 2021-12-17 DOI: 10.17504/protocols.io.b2yvqfw6
    AI 解读
    浏览 1676

    Disclaimer

    DISCLAIMER – FOR INFORMATIONAL PURPOSES ONLY; USE AT YOUR OWN RISK

    The protocol content here is for informational purposes only and does not constitute legal, medical, clinical, or safety advice, or otherwise; content added to protocols.io is not peer reviewed and may not have undergone a formal approval of any kind. Information presented in this protocol should not substitute for independent professional judgment, advice, diagnosis, or treatment. Any action you take or refrain from taking using or relying upon the information presented here is strictly at your own risk. You agree that neither the Company nor any of the authors, contributors, administrators, or anyone else associated with protocols.io, can be held responsible for your use of the information contained in or linked to this protocol or any of our Sites/Apps and Services.

    Abstract

    This protocol presents the procedures to estimate the concentration of cells' suspensions using an hemocytometer

    Steps

    1.

    Wash the hemocytometer with a washer bottle and dry it using a paper towel

    2.

    Cover the hemocytometer with the coverslip and place it on a flat surface

    3.

    Vortex briefly the cell suspension whose concentration will be calculated

    4.

    Using a micropipette collect 10-100 µl of the previously agitated suspension

    5.

    Place the micropipette tip on the hemocytometer edge near the coverslip

    6.

    Empty the spore suspension slowly into the chamber of the hemocytometer

    7.

    Place the hemocytometer in the microscope's mechanical stage and secure it using the stage clip

    8.

    Turn on the microscope and focus on the hemocytometer

    9.

    Look out for the first square to be counted

    10.

    Count the number of cells on each square

    Note
    If the cells are touching the upper or the left lines of the square, cells are to be counted, but if they are touching the lower or right line they should not be counted

    11.

    Count all the squares following a zig-zag

    Citation

    12.

    Write down the number of cells on each square

    Note
    The lower limit for accurate counting of cells in a hemocytometer is usually considered to be 2.5 x 105/ml. For cell numbers greater than 2.5 x 106/ml, it is generally recommended that the sample be diluted.

    Note
    Count the number of cells on at least five squares. Ideally, repeat the procedure twice for higher accuracy

    13.

    Estimation of the concentration

    Note
    According to the cell's size cells can be counted on squares 1, 2 or 3

    13.1.

    If cells were counted on squares of size 1

    Concentration (cells/ml)= Number of counted cells*10,000/Number of squares counted

    Note
    If the cells' suspension has been diluted the formula should be adjusted to: Concentration (cells/ml)= Number of counted cells10,000/Number of squares counteddilution

    13.2.

    If cells were counted on squares of size 2

    Concentration (cells/ml)= Number of counted cells*160,000/Number of squares counted

    Note
    If the cells' suspension has been diluted the formula should be adjusted to:Concentration (cells/ml)= Number of counted cells160,000/Number of squares counteddilution

    13.3.

    If cells were counted on squares of size 3

    Concentration (cells/ml)= Number of counted cells*250,000/Number of squares counted

    Note
    This square size is typically used to count Plasmodiophora brassicae resting spores

    Note
    If the cells' suspension has been diluted the formula should be adjusted to:Concentration (cells/ml)= Number of counted cells250,000/Number of squares counteddilution